Embryos of seven patients did not progress as expected and were not suitable for embryo transfer

Status: 
Ready to upload
Record number: 
1926
Adverse Occurrence type: 
MPHO Type: 
Estimated frequency: 
Never reported before
Time to detection: 
N/A
Alerting signals, symptoms, evidence of occurrence: 
Laboratory staff noticed that eggs and embryos of seven patients being cultured and stored in a specific incubator had failed to progress.
Demonstration of imputability or root cause: 
Equipment failure: incubator’s oxygen sensor was faulty, leading not to detect a high concentration of nitrogen and then, resulting a low oxigen concentration ( this should be 5% but was found to be 3%)
Imputability grade: 
2 Probable
Groups audience: 
Suggest new keywords: 
embryos
equipment failure
incubator
oxigen
laboratory
Suggest references: 
HFEA Adverse incidents in fertility clinics lessons to learn 2013