Loss of donor sperm belonging to 250 patients

Status: 
Ready to upload
Record number: 
1925
Adverse Occurrence type: 
MPHO Type: 
Estimated frequency: 
Never reported before
Time to detection: 
N/A
Alerting signals, symptoms, evidence of occurrence: 
The liquid nitrogen in the tank run out earlier than expected, leading to loss of 250 donor sperm samples.
Demonstration of imputability or root cause: 
The investigation revealed that a valve supplying liquid nitrogen to the storage tank containing the samples had been left closed, preventing liquid nitrogen from entering the system. There was also a damaged connection hose, resulting in the liquid nitrogen evaporating in the hose and emptying the entire supply tank while not filling the storage tank.
Imputability grade: 
3 Definite/Certain/Proven
Groups audience: 
Suggest new keywords: 
liquid nitrogen
supply
tank
sperm
frozen
semen
loss
Suggest references: 
HFEA Adverse incidents in fertility clinics lessons to learn 2013
Expert comments for publication: 